• Nano Laser Particle Size Analyzer
  • Nano Laser Particle Size Analyzer
  • Nano Laser Particle Size Analyzer
Nano Laser Particle Size Analyzer
  • ST-NP900
  • 1-10000nm
  • 0.1mg/mL ~ 100mg/mL
  • ≤1%
  • ≤1%

Nano Laser Particle Size Analyzer is mainly designed to the detection of particle size and distribution of nanomaterials, nano emulsions, nano powders, nano borides, nano non-metals, coatings, dyes, pigments, pharmaceuticals, proteins, nano calcium carbonate, photosensitive materials, additives, graphite, exosomes, and other nanoparticles in fields such as new materials, pharmaceuticals, bioengine

Product Description

The photon related nano particle sizer is the product of the national science and technology small and medium-sized enterprise technology innovation fund project. It is also the first batch of nano particle sizers in China that adopt the principle of dynamic light scattering. The instrument adopts the principle of dynamic light scattering and photon correlation spectroscopy. Due to the Brownian motion of particles in the suspension, the light intensity fluctuates with time. The micro nano digital correlator technology is used to process the pulse signal to obtain the diffusion information of particle motion. The particle size and its distribution are calculated by using Stokes Einstein equation.

It is equipped with a dual wavelength laser. It can effectively detect some samples that have light absorption properties and cannot be detected by traditional single wavelength lasers. The nano laser particle sizer focuses on the particle size distribution detection of colored particles, such as pigments and dyes.



Product advantages

1. High sensitivity and SNR

Based on the principle of dynamic light scattering, the Hamamatsu photomultiplier tube with high sensitivity and strong signal-to-noise ratio and the independently developed cr256 digital correlator can complete the collection and function calculation of dynamic scattered light in real time without damaging or interfering with the nano particle system, effectively calculate the particle size, and ensure the accuracy of the test results.

2. Strong anti-interference

The optical path system built with optical fiber technology makes the photon correlation spectrum detection system not only small in size, but also has strong anti-interference ability, ensuring the stability of the test results.

3. Reliable temperature control system

The semiconductor temperature control system is adopted, and the temperature fluctuation is controlled at ± 0.1 ℃, which ensures that the sample is always in a constant temperature state during the test process, and avoids the test deviation caused by the change of liquid viscosity and Brownian motion speed caused by temperature change, so as to ensure the reliability of the test results.

4. Dual wavelength laser, intelligent switching

Dual wavelength λ= 532 nm λ= 405 nm) laser, which can be switched intelligently, and can effectively detect some samples with light absorption properties that cannot be detected by a single wavelength laser.



Application

It is mainly designed to the detection of particle size and distribution of nanomaterials, nano emulsions, nano powders, nano borides, nano non-metals, coatings, dyes, pigments, pharmaceuticals, proteins, nano calcium carbonate, photosensitive materials, additives, graphite, exosomes, and other nanoparticles in fields such as new materials, pharmaceuticals, bioengineering, life sciences, chemicals, coating, and food.



Technical Data

Product Model

ST-NP900

Compliance Standard

GB/T29022-2012/ISO22412:2008

Measuring Range

1-10000nm

Measuring Concentration

0.1mg/mL ~ 100mg/mL

Accuracy Error

≤1% (CRM D50)

Repeatability Error

≤1% (CRM D50)

Laser

Main Light Source: Semiconductor laser, λ=532nm; Power: 1-40mW (adjustable)

Auxiliary Light Source: Blue light semiconductor laser, λ=405nm; Power: >2mW

Detector

Photomultiplier Tube (PMT)

Scattering Angle

90°

Sample Cell

10×10×40mm (1-4mL)

Measuring Temperature

5~90℃ (minimum precision 0.1℃)

Temperature Control Accuracy

±0.1℃

Measuring Speed

<5min

Digital Correlator Model

CR512

Auto-correlation Channels

512

Base Channels

4

Physical Channels

10000

Unit Delay Time

100ns~10ms

Operation Mode

SOP setting supported, fully automatic test, one-click automatic test, one-click switch between Chinese and English

Statistical Comparison

Supports comparison of multiple test results; users can compare and analyze test data of different sample batches, before/after sample processing, or at different processing stages, to guide product quality and cost control

Data Processing

Intelligent statistical processing of test data, outputting the most representative test results

Custom Analysis

Customizable output of any characteristic particle size between D1~D100, cumulative percentage of particle size above/below a set value, and cumulative percentage of a specified particle size interval, to meet different industry requirements for particle size characterization

Result Output

Typical particle size values, interval particle size distribution (differential distribution) data & graphs (histogram/curve), cumulative particle size distribution (integral distribution) data & curves, etc.

Test Report Export

Test reports can be exported to Word, Excel, Image (.bmp), PDF and other formats, meeting the needs of report viewing, data reprocessing and academic citation

Language Support

Chinese-English bilingual interface; additional language interfaces can be customized according to user requirements

GMP Compliance

GMP module (optional)

Overall Dimensions

600×380×230mm

Net Weight

12Kg

Mail consultation