• Single Beam Atomic Absorption Spectrophotometer for Laboratory AAS-3000
Single Beam Atomic Absorption Spectrophotometer for Laboratory AAS-3000
  • 170-900nm

This high-performance Atomic Absorption Spectrophotometer AAS-3000 is designed for elemental analysis in environmental, food, agricultural, metallurgical, and pharmaceutical applications. Built with advanced optics, a reliable atomization system, and intelligent software, it delivers exceptional sensitivity, stability, and ease of use for flame atomic absorption analysis.

Product Description

Product Introduction

This high-performance Atomic Absorption Spectrophotometer AAS-3000 is designed for elemental analysis in environmental, food, agricultural, metallurgical, and pharmaceutical applications. Built with advanced optics, a reliable atomization system, and intelligent software, it delivers exceptional sensitivity, stability, and ease of use for flame atomic absorption analysis.


Product Features

1.     Light Source System

  • Automatic 8-lamp turret with automatic optimization and alignment, supporting 2 positions for high-performance lamps.
  • Simultaneous 8-lamp power supply with preheating capability, ensuring stable light output.
  • Freely editable lamp position memory function, eliminating the need for coded lamps.

Optical System

  • Czerny-Turner monochromator with a high-resolution diffraction grating (1800 lines/mm).
  • 7 selectable spectral bandwidths, with automatic setting and optimization.
  • Rigid cast optical platform with ultra-low deformation under high/low temperature conditions, ensuring long-term stability.
  • Dual-beam optics to effectively suppress baseline drift, reduce warm-up time, and improve accuracy.
  • Excellent baseline stability: Drift better than 0.001 Abs over 8 hours of continuous operation.

2.     Atomization System

  • Constant & Transverse Magnetic Field Zeeman Background Correction: Delivers full-element, full-wavelength background correction, effectively eliminating background interference up to 2 Abs.
  • Enhanced magnetic field intensity (up to 1.0T) improves analytical sensitivity.
  • Automatic ignition, gas flow precision control, and intelligent flame height/flow self-adaption.
  • Multi-level safety protection: Alarm for gas leakage, abnormal flow, insufficient air pressure, and flame extinction; equipped with an emergency flame shutoff button.
  • Cooling water and temperature monitoring for magnetic field protection; foot-operated reading function for hands-free operation.

3.     Software & Communication

  • Intelligent control software compatible with Windows 7 & 10.
  • One-key instrument optimization and multi-task analysis support.
  • Automatic curve fitting, re-sloping, and concentration calculation functions.
  • Ethernet interface for stable and reliable data communication.


Technical Data

Parameter

Specification

Wavelength Range

170 nm ~ 900 nm

Wavelength Accuracy

≤ ±0.10 nm

Wavelength Repeatability

≤ 0.05 nm

Baseline Stability

60 min baseline drift ≤ 0.0005 Abs; instantaneous noise ≤ 0.0005 Abs

Dynamic Baseline Stability

30 min baseline drift ≤ 0.001 Abs; instantaneous noise ≤ 0.001 Abs

Resolution

Spectral bandwidth deviation ≤ 0.02 nm; valley-peak energy ratio ≤ 25% (Mn 279.5 nm / 279.8 nm)

Cu Detection Performance

Detection limit ≤ 0.002 µg/mL; characteristic concentration ≤ 0.03 µg/mL/1%; precision ≤ 0.25%

Background Correction

Better than 150 times

Dimensions (L×W×H)

1010 × 620 × 630 mm

Net Weight

115 kg


Flame Analysis Performance Data (Representative Elements)

Element

Wavelength (nm)

Flame Type

Detection Limit (µg/mL)

Characteristic Concentration (µg/mL)

Recommended Range (µg/mL)

Minimum Detectable Concentration (µg/mL)

Ag

328.1

Oxidizing

0.0028

0.023

0.1–3.5

0.05

Cu

324.8

Oxidizing

0.0043

0.024

0.1–5.0

0.03

Fe

248.3

Oxidizing

0.0183

0.440

0.3–6.0

0.15

Pb

283.3

Oxidizing

0.0056

0.137

0.5–20

0.25

Zn

213.9

Oxidizing

0.0014

0.007

0.03–1.0

0.015

(Full data for 32 elements available upon request)


Extended Accessories

1. Flame Autosampler

  • Patented quick-release design for easy reagent/sample vessel replacement.
  • Capacity: 70 vessels (15 for reagents, 55 for samples), vessel volume ≥ 20 mL.
  • Customizable vessel position; dimensions: 450 × 300 × 450 mm (L×W×H), 14 kg.

2. Specialized Nebulizers

  • Organic phase-resistant nebulizer, HF acid-resistant nebulizer, etc., for challenging matrices.

3. Audit Trail Software

  • Compliant with FDA 21 CFR Part 11 regulations, ideal for pharmaceutical and regulated industries.
Mail consultation
Please feel free to give your inquiry in the form below.We will reply you in 24 hours.